6 Cochlear™ MRI Kit User Guide © Cochlear Limited, 2022 7
Intended purpose
The Cochlear MRI Kit is intended to prevent the dislodgement of implanted
magnets in a hearing implant during a Magnetic Resonance Imaging (MRI)
procedure.
The Cochlear MRI Kit is for single-use only.
Indications
The Cochlear MRI Kit is indicated for hearing implant recipients who require
an MR scan at 1.5 T and have been assessed by medical professionals as
suitable for an MR scan.
The Cochlear MRI Kit is indicated for a recipient with the following
compatible Osia implants:
Osia OSI100 implant
Osia OSI200 implant
CAUTION
The Cochlear Osia implants will create shadowing on the MR image
near the implant, resulting in a loss of diagnostic information.
Refer to the Image interference and artefacts section of the
Cochlear Osia MRI Guidelines.
Contraindications
The Cochlear MRI Kit is contraindicated for use with MR scans other than
1.5 T and for recipients with any implant except those identified in the
indications.
For information on identifying Cochlear Osia implants without surgical
intervention, refer to the Cochlear Osia MRI Guidelines.