Explanatory notes:
a. Field strengths from fixed transmitters, such as base
stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the
location in which the processor is used exceeds the
applicable RF compliance level above, the processor
should be observed to verify normal operation. If
abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the
processor.
b. Over the frequency range 150 kHz to 80 MHz, field
strengths should be less than 3 V/m.
Recommended separation distances
Your processor is intended for use in an electromagnetic
environment where the radiated RF disturbances are
controlled.
To prevent electromagnetic interference, maintain a
minimum distance between the portable and mobile RF
communications equipment (transmitters) and the device as
recommended below, according to the maximum output
power of the communications equipment.