Product Manual
Page 43
No.
Action
Image guided instruction
8. 7
IMMEDIATELY after
collecting the patient sample,
place the tip of the sample
transfer tube at a 45º angle
into the sample well on the
test strip in front of the wheel
where you see the flashing
green light. Gently touch the
tip down onto the sample
well. Depress plunger
completely to apply blood
sample.
Discard the sample transfer
tube in a biohazard waste
container.
9. 8
When the blood sample is
detected, the meter displays
a ‘Testing Please Wait’
message.
Note: If this screen is not displayed,
either not enough blood sample was
applied, or the sample had air bubbles in
it. Remove the strip and retest with a new
strip and fresh fingerstick.
10. 9
When testing is complete,
the meter beeps (if sound is
turned ON). The results (INR
and PT in seconds) are
displayed on the screen
along with date and time of
the test.
Note: Memo field allows user to make
notes along with the results.
Upon clicking the Check mark icon, the
main screen is displayed.
Note: Refer to the “Troubleshooting”
section of this manual if the meter
displays messaging, like for example:
CLOT TIME TOO SHORT or NO CLOT
DETECTED.