Product Manual

Page 42
No.
Action
Image guided instruction
respective fields. Press the
forward button.
Note: Make sure the expiration date on
the strip packaging has not passed.
Contact your Coag-Sense distributor for
help with reordering.
3. 3
Open the packaging of the
test strip by tearing the
notched end.
4. 4
Holding the round end,
gently push the strip
completely into the meter.
The strip fits snuggly when
pushed all the way toward
the back wall of the strip
insertion area.
5. 5
The meter warms the strip
(for 25 seconds) to operating
temperature. The display
shows a countdown in
seconds.
Note: Do not apply test sample until the
warm-up is complete and the meter
display shows ‘Apply Sample’.
6. 6
While the meter is warming up, get ready to perform a
fingerstick. See “Collecting a Fingerstick Sample” in this
manual.
7. 5
When the warm-up is
complete, the meter beeps (if
sound is turned ON) and the
screen displays a ‘Apply
Sample’ message.
Note: You now have up to 2 ½ minutes
to perform a fingerstick and apply the
sample to the test strip.