Product Manual

Page 38
No.
Action
Image guided instruction
11.
When the sample is
detected, the meter displays
a ‘Testing Please Wait’
message.
Note: If this screen is not displayed
within 8 seconds not enough blood
sample was applied. DO NOT attempt to
add more sample. Stop the test and
retest with a new strip and fingerstick.
12.
When testing is complete,
the meter beeps (if sound is
turned ON). The results (INR
and PT seconds) are
displayed on the screen
along with the date and time
of the test.
Note: Memo field allows user to make
notes along with the results.
Upon clicking the Check mark icon, the
main screen is displayed.
Note: Refer to the “Troubleshooting”
section of this manual if the meter
displays messaging, like for example:
CLOT TIME TOO SHORT or NO CLOT
DETECTED.
13.
Remove the test strip and
dispose in a biohazard
collection container.
Note: Repeat the test if the results
seem unusually low or high. If the
results still seem unusual after a second
test, contact Technical Support.
14.
User may export recorded
results into the electronic
medical record system by
connecting to a computer or
middleware system.
User may also print the
results as well, refer to the