10. Performing a PT Test
WARNING: Place the meter on a stationary, level
surface for testing. DO NOT move the meter or allow it
to vibrate during the test. Unreliable results may occur.
Wear gloves and follow all applicable hygiene and
safety procedures.
Follow the below steps to perform a patient test:
No.
Action
Image guided instruction
1.
Turn the meter ON by
pressing and holding the
(POWER) button on the right
side of the meter.
Note: The message field on the first
screen will display errors encountered
during self-check if any.
2.
Press the Login icon.
Operator is now directed to
enter login credentials or
scan operator barcode using
optional barcode scanner.
3.
Successful login directs the
user to the Home screen.
Press the Test icon on
display screen.
Note: If the control test fails, patient test
feature will be disabled, please contact
CoaguSense technical support for
assistance.