Product Manual

Page 31
No.
Action
Image guided instruction
12. 1
3
When the control activation
solution is properly applied
and detected, the flashing
green light will turn off, and
the meter displays ‘Testing
Please Wait’.
Note: If this screen is not displayed
within 8 seconds not enough solution
was applied. Remove the strip. Retest
with a new control strip. DO NOT
attempt to add more solution to the
strip.
13. 1
4
When testing is complete,
the Pass/Fail results are
displayed in PT units. Date
and Time are also
displayed.
Note: Control test results only display
PT seconds, this is to avoid confusing
control strip INR results with patient
test strip INR results.
14. 1
5
Repeat Steps 3-13 for ‘High
control strip.
15. 1
6
Once the controls have
been successfully tested,
remember to remove and
discard the control strips.
You can now proceed to
testing patient blood
samples. If you are not
going to test, turn off the
meter by pressing and
holding the POWER
button. The opened control
activation solution may be
used until the expiration
date.
Note: If control test fails, repeat the test with a new strip. If the control test
continues to FAIL, please contact your POC testing administrator or Coag-Sense
Technical Support for assistance.