Product Manual

Page 30
No.
Action
Image guided instruction
9. 9
Insert the black plastic
plunger into the end of the
glass capillary tube with the
red stripe. Use care to
avoid hitting white plug.
(Preloaded tubes are also
available).
10. 1
0
Open the control activation
solution and hold at an
angle to allow insertion of
the transfer tube. Insert
transfer tube into control
activation solution. Let
capillary action fill until
solution flow stops at white
plug.
11. 1
2
The meter beeps once and
displays “Apply Control
Solution” when it is ready
for the control strip
activation solution.
Note: You now have up to 2 ½ minutes
to apply the activation solution to the
control strip.
Rest hand on the meter or
counter top to steady.
Insert transfer tube tip into
sample application well of
test strip, touching tip down
at flashing green light in
front of wheel. Depress
black plunger completely to
dispense the activation
solution.