Product Manual

Page 24
No.
Action
Image guided instruction
External Connection
8.
a) Screen lists server IP
address for socket
communication
b) Open/Close port for serial
communication with external
program
c) Connects PT/INR meter to
PC program using Bypass
Mode.
Admin Settings
9.
QC Lock Settings:
If a Control test (High Control
or Low control) does not
produce a result, or if the
result is outside of the target
value range, the meter can be
locked from further patient
testing. The frequency for
required testing can be set.
Note: Refer to QC Lockout section below
for detailed information.
10.
Sleep Settings:
If left unattended, the meter
powers itself off. You manage
the time in minutes before
going into sleep mode or turn
off the sleep function so the
meter will always be on until
manually powered OFF.