Directions for Use
7. Meter Setup
Operating Conditions
To ensure that your Coag-Sense PT/INR system is working correctly, be
sure the following conditions are met:
• Be sure that the meter and strips are at room temperature
before use. Operating conditions are between 65ºF and
90ºF (18ºC and 32ºC). The meter will not allow a test
to proceed until the meter is at room temperature.
• Relative humidity should be between 10% and 85%,
without condensation, for testing.
• Avoid dropping the meter or treating it roughly.
• Use the meter only on a level, stable surface.
• Do not move or touch the meter during testing.
Power ON/OFF
• The Coag-Sense PT/INR Professional System can be
operated with the power adapter provided. The power
adapter also serves as a charger.
• Place the meter on a flat, stable surface. To turn the
meter ON/OFF, press and hold the POWER button of
the right side of the meter.