Product Manual
Page 13
If any alcohol remains (or is re-introduced) on the finger, it may
cause unreliable results.
• Do not use wipes containing chlorhexidine gluconate, as it may
produce unreliable results.
• The quality of fingerstick and the sample delivery technique
are important to the test results. If there is a question about the
sample or sample collection, obtain a new strip, repeat the
fingerstick on a different finger, and test again.
• If you need to repeat a test, use a different finger for the fingerstick,
since blood may have started to clot on the first finger, which may
cause unreliable results.
• If there is a bubble or an air pocket showing in the blood sample in
the collection tube, start the test over. Use a new fingerstick (using
a different finger and collection tube) or results may be unreliable.
Meter
• The meter has a built-in rechargeable lithium polymer battery (3.7V,
2350 mAh).
• Use only the included Coag-Sense Power adapter with the meter or
damage to the meter may result.
• The meter is a delicate instrument, and should be handled with care.
Dropping or other mishandling may cause damage to the meter. If
this should occur, call Technical Support.
• Do not allow any liquids to spill on the meter. If this should occur,
call Technical Support.
• Do not put the meter in liquid. Do not allow liquids to get into any of
the connectors or plugs on the meter.
• Only use the method provided in this User Manual to clean the
Coag-Sense PT/INR meter.
• Do not move or touch the meter while it is running a test. Unreliable
results may occur.