Product Manual

Page 13
If any alcohol remains (or is re-introduced) on the finger, it may
cause unreliable results.
Do not use wipes containing chlorhexidine gluconate, as it may
produce unreliable results.
The quality of fingerstick and the sample delivery technique
are important to the test results. If there is a question about the
sample or sample collection, obtain a new strip, repeat the
fingerstick on a different finger, and test again.
If you need to repeat a test, use a different finger for the fingerstick,
since blood may have started to clot on the first finger, which may
cause unreliable results.
If there is a bubble or an air pocket showing in the blood sample in
the collection tube, start the test over. Use a new fingerstick (using
a different finger and collection tube) or results may be unreliable.
Meter
The meter has a built-in rechargeable lithium polymer battery (3.7V,
2350 mAh).
Use only the included Coag-Sense Power adapter with the meter or
damage to the meter may result.
The meter is a delicate instrument, and should be handled with care.
Dropping or other mishandling may cause damage to the meter. If
this should occur, call Technical Support.
Do not allow any liquids to spill on the meter. If this should occur,
call Technical Support.
Do not put the meter in liquid. Do not allow liquids to get into any of
the connectors or plugs on the meter.
Only use the method provided in this User Manual to clean the
Coag-Sense PT/INR meter.
Do not move or touch the meter while it is running a test. Unreliable
results may occur.