Product Manual

Page 38
No
Action
Image guided instruction
9.
When testing is complete,
the meter beeps (if sound
is turned ON). The results
(INR and PT seconds) are
displayed on the screen
along with date and time
of the test.
Note: Memo field allows you to make
notes such as medication or diet
change along with the results. Upon
clicking the Check mark icon, the
main screen is displayed.
Note: Refer to the
“Troubleshooting” section of this
manual if the meter displays
messaging, for example:
CLOT TIME TOO SHORT or NO
CLOT DETECTED.
10.
Remove the test strip and
properly dispose.
Note: Repeat the test if the results
seem unusually low or high. If the
results still seem unusual after a
second test, contact Technical
Support.
11.
You may print the results if
you purchased the
optional portable printer.
Refer to the “Printing”
section in this manual.
Note: The meter stores about 2000
patient test results in memory with
the time and date stamp. Refer to
“Reviewing the Memory” in this
manual for more information.
12.
Turn the meter OFF by
pressing and holding the
POWER button when
you are finished testing. If
left unattended for a set
time, (User preferred