Product Manual
Page 37
No
Action
Image guided instruction
While the meter is warming up, get ready to perform a fingerstick. See
“Collecting a Fingerstick Sample” section in this manual.
When the warm-up is
complete, the meter
beeps (if sound is turned
ON) the screen displays
‘Apply Sample’ message.
Note: You now have up to 2 ½
minutes to perform a fingerstick
and apply the sample to the test
strip.
7.
IMMEDIATELY after
collecting your blood
sample, place the tip of
the sample transfer tube
at a 45º angle into the
sample well on the test
strip in front of the wheel
where you see the
flashing green light.
Gently touch the tip down
onto the sample well.
Depress the plunger
completely to dispense
blood sample.
Note: Depress black plunger
completely to dispense the blood.
8.
When the sample is
detected, the meter
displays a ‘Testing Please
Wait’ message.
Note: If this screen is not displayed
within 8 seconds not enough blood
sample was applied. DO NOT
attempt to add more sample. Stop
the test and retest with a new strip
and fingerstick.