Product Manual

Page 36
No
Action
Image guided instruction
If NFC tag information is
not available, you may
manually enter the Lot #
and Barcode # using the
keypad on the
touchscreen into the
respective fields.
Press the forward button.
Note: Make sure the expiration date
on the strip packaging has not
passed. Contact your Coag-Sense
distributor for help with reordering.
4.
Open the packaging of the
test strip by tearing the
notched end.
5.
Holding the round end,
gently push the strip
completely into the meter.
The strip fits snuggly
when pushed all the way
toward the back wall of the
strip insertion area.
6.
The meter warms the strip
(for 25 seconds) to
operating temperature.
The display shows a
countdown in seconds.
Note: Do not apply test sample
until the warm-up is complete and
the meter display shows Apply
sample’.
Patient ID is not displayed on personal meter version