10. Performing a PT Test
WARNING: Place the meter on a stationary, level surface for
testing. DO NOT move the meter or allow it to vibrate during
the test. Unreliable results may occur.
Follow the steps below to perform a patient test:
No
Action
Image guided instruction
1.
Turn the meter ON by
pressing and holding the
(POWER) button on the
right side of the meter.
After startup and self-
check, the display will
show the Home Screen.
2.
To perform a Patient test,
press the Test icon on
display screen.
3.
Patient Strip lot
confirmation screen
displays the Lot
information of the strip
that was last recorded.
Proceed with testing if the
test strip is from the same
lot.
Otherwise, touch the NFC
tag against the NFC tag
scanner on the meter, the
Lot # (six-digit numeric
identifier) and Barcode #
(eight-digit numeric
identifier) will auto
populate.