Product Manual

Page 33
No.
Action
Image guided instruction
4.
Remove the cap from the
single use lancet. Place it
against the skin. Holding
the body of the lancet,
push down firmly against
the finger to lance the
surface of the skin. Do
not lance finger until
meter displays “APPLY
SAMPLE”. A minimum of
10µl of collected blood
sample is required.
Note: The blood should flow freely.
If it doesn’t, gently squeeze the
finger to get it started. Lowering the
patient’s hand and arm so that the
fingertip is below the heart helps the
blood drop form.
5.
When ready to collect the
drop of blood, hold the
Sample Transfer Tube
horizontal. Touch tip to
bead of blood and let
capillary action fill until
blood flow stops at white
plug. Squeeze finger to
generate additional blood
if required to completely
fill to white plug.
Note: Blood must reach the white
plug
WARNING: Squeezing the fingerstick site excessively
(milking releases interstitial “tissue layer” fluid that cause
unreliable results.