Product Manual

Page 29
No.
Action
Image guided instruction
solution flow stops at
white plug.
9.
The meter beeps once
and displays “Apply
Control Solution” when it
is ready for the control
strip activation solution.
Note: You now have up to 2 ½
minutes to apply the activation
solution to the control strip.
Rest hand on meter or
counter top to steady.
Insert transfer tube tip
into sample application
well of test strip, touching
tip down at flashing green
light in front of wheel.
Depress black plunger
completely to dispense
the activation solution.
10.
When the control
activation solution is
properly applied and
detected, the flashing
green light will turn off,
and the meter will display
‘Testing Please Wait’.
Note: If this screen is not displayed
within 8 seconds not enough
solution was applied. Remove the
strip. Retest with a new control strip.
DO NOT attempt to add more
solution to the strip.