Product Manual

Page 28
No.
Action
Image guided instruction
4.
Open the packaging of
the selected control strip
by tearing the notched
end.
5.
Holding the round end,
gently push the strip
completely into the meter.
The strip fits snuggly
when pushed all the way
toward the back wall of
the strip insertion area.
6.
The meter warms the
strip (for 25 seconds) to
operating temperature,
display shows a
countdown in seconds.
Note: Do not apply the control
activation solution until the warm-up
is complete and the meter display
shows ‘Apply Control Solution’.
7.
Insert the black plastic
plunger into the end of the
glass capillary tube with
the red stripe. Use care to
avoid hitting white plug.
Note: Your home testing service
provider may have supplied you with
transfer tubes that have the
plungers pre-loaded.
8.
Open the control
activation solution and
hold at an angle to allow
insertion of the sample
transfer tube. Insert
transfer tube into control
activation solution. Let
capillary action fill until