Product Manual
Page 26
8. Performing a Control Test
There are 2 low control strips, 2 high control strips and a control strip
activation solution shipped with each Coag-Sense test strip kit.
Controls should be tested immediately upon receipt of each new lot
number. Extra controls may be ordered separately if more frequent QC
testing is required. Please note that your home testing service provider
may not include control strips in your supply shipment.
Note: The following directions are for running a low control strip. When
this procedure is complete, run a high control strip. The controls may
be run in any order. The meter will display and store the results in PT
seconds only. The meter does not use or require results from the
control strips prior to running a patient test strip.
Note: If an error message appears, consult the ‘Troubleshooting’
section of this manual.
WARNING: DO NOT move or touch the meter while it is
running a test. Unreliable results may occur.
Follow these steps to perform a test on a low or high control.
No.
Action
Image guided instruction
1.
Turn the meter ON by
pressing and holding the
(POWER) button on
the right side of the meter.
After the system starts
the Home Screen will be
displayed.
2.
To perform a control test,
press the Control icon on
display screen.
Select from the following
Low Control Test or High
Control Test.
Note: The sequence of running the
controls is not important.