7. Meter Setup
Operating Conditions
To ensure that your Coag-Sense PT/INR system is working correctly, be
sure the following conditions are met:
• Be sure that the meter and strips are at room temperature before use.
Operating conditions are between 65ºF and 90ºF (18ºC and 32ºC).
The meter will not allow a test to proceed until the meter is at
room temperature.
• Relative humidity should be between 10% and 85%, without
condensation, for testing.
• Avoid dropping the meter or treating it roughly.
• Use the meter only on a level, stable surface.
• Do not move or touch the meter during testing.
Power ON/OFF
• The Coag-Sense PT/INR Patient Self-Test System can be operated
with the power adapter provided. The power adapter also serves as
a charger.
• Place the meter on a flat, stable surface. To turn the meter ON/OFF,
press and hold the POWER button of the right side of the meter.