Product Manual

Page 15
CoaguSense, you are responsible for meeting the requirements of
this standard.
In the unlikely event of an electric power surge (i.e., severe static
discharge during a thunderstorm), when using the power adapter, the
display screen may go blank. If this occurs, unplug the power supply
from the back of your meter, wait 5 seconds and plug it back in.
Normal operation should return, but you may have to reset the time
and date.
DO NOT OPEN THE METER. Do not attempt to repair or modify this
meter. The Coag-Sense meter does not require any periodic
maintenance and there are no user serviceable parts inside. If you
have problems, please contact Technical Support. The Coag-Sense
Prothrombin Time (PT)/INR Monitoring System needs special
precautions regarding EMC and needs to be put into service
according to the EMC information provided in this manual.”
“Portable and mobile RF communications equipment can affect The
Coag-Sense Prothrombin Time (PT)/INR Monitoring System.”
“The use of accessories, transducers and cables other than those
specified by CoaguSense, may result in increased EMISSIONS or
decreased IMMUNITY of the EQUIPMENT.”
“The Coag-Sense Prothrombin Time (PT)/INR Monitoring System
should not be used adjacent to or stacked with other equipment and
that if adjacent or stacked use is necessary, the Coag-Sense
Prothrombin Time (PT)/INR Monitoring System should be observed
to verify normal operation in the configuration in which it will be used.”
Test Strips/ Control Strips/Control Strip Activating Solution
The test strips are designed for single use only. Do not reuse the test
strips.
Patient samples, used strips, transfer tubes and lancets are
potentially infectious. Discard used materials in a puncture resistant,
biohazard waste container using universal precautions.