Product Manual

Page 13
Use only fresh fingerstick capillary blood for testing. The blood
should only come in contact with the products provided with the
Coag-Sense PT/INR system. Other products may have anti-
coagulant agents on their surfaces and result in unreliable test
results.
Squeezing the fingerstick site excessively (milking) releases
interstitial “tissue layer” fluid that can cause unreliable results.
The fingerstick site can be washed with warm water and soap, and
then completely dried. The site must be clean of all hand oils/lotions
and foreign matter, which may cause unreliable results.
If Isopropyl Alcohol (IPA) wipes are used, wipe the fingerstick
site with a gauze pad and make sure the site is completely dry.
If any alcohol remains (or is re-introduced) on the finger, it may
cause unreliable results.
The quality of fingerstick and the sample delivery technique
are important to the test results. If there is a question about the
sample or sample collection, obtain a new strip, repeat the
fingerstick on a different finger, and test again.
If you need to repeat a test, use a different finger for the fingerstick,
since blood may have started to clot on the first finger, which may
cause unreliable results.
If there is a bubble or an air pocket showing in the blood sample in
the collection tube, start the test over. Use a new fingerstick (using
a different finger and collection tube) or results may be unreliable.
Meter
The meter has a built-in rechargeable lithium polymer battery (3.7V,
2350mAh).
Use only the included Coag-Sense Power adapter with the meter
or damage to the meter may result.