5. Warnings and Precautions
Patients taking Warfarin (Coumadin) and other oral
blood thinners should consult with their healthcare
provider before adjusting their dosage.
Patients should consult with their doctor for their
appropriate INR therapeutic range.
Patients who have recently taken or are currently taking
any type of Heparin or Low Molecular Weight Heparin
anticoagulant should not use this test system and should
consult their doctor.
The system should also not be used to monitor patients
on direct oral anticoagulants (DOACs) including Factor
Xa and Direct Thrombin inhibitors.
Test Site and Blood Sample
The Coag-Sense PT/INR system is for in vitro diagnostic use only.
The Coag-Sense meter will not produce a result if the test strip is
past its expiration date.
The quality of the blood sample can affect PT test results. A blood
sample of poor quality can produce unreliable results. Read the
section on Collecting a Fingerstick Sample for more
information.
Blood samples must be applied to the test strip immediately after
collection or the blood begins to clot, causing unreliable results.
The blood sample transferred to the test strip must be a minimum
of 10 µL in volume. Low sample volume may cause an error
message.