Product Manual
Page 10
3. Meter Specifications
Operating Temperature
65°F to 90°F (18°C to 32°C)
Operating Humidity
10% to 85% (without condensation)
Storage Temperature
32°F to 122°F (0°C to 50°C)
Storage Humidity
20% to 80%
Memory
Capable of storing up to,
• 2000 patient test results with date and time
• 500 control test results with date and time
• 1000 Operator accounts
Lithium Battery
Rechargeable Lithium Polymer Battery
(3.7V, 2350mAh)
Battery Capacity
Fully charged (6 hours of charging) can
run ~200 tests
Power Input
120V AC Adapter (Use with Coag-Sense
Adapter Only)
Power Output
5.0V, 2.0A
Blood Sample Size
10-12 µL
Communication Port
Micro and Standard USB
Size in mm
(Height x Width x Depth)
152 x 100 x 29.5
Weight in g
315g
Equipment Classification
Class II with external power supply.
Internally powered when operated with
battery. IPX0 rating.
WARNING: Use the Coag-Sense meter along with the
provided Power Adapter only.