Service manual
UCW and Ultraview 1700 Monitors — Service Manual
4-6
Verifying Monitor Functions
To verify that the monitor functions correctly with parameter modules:
1. Insert a module that includes ECG without the patient cables connected.
• Verify that the vertical ECG parameter key appears, a heart rate of ??? is displayed, and
the message LEADS OFF is in the waveform zone.
2. Connect a waveform simulator to the ECG input with a 5-lead patient cable, and set the
simulator to a known rate.
• Verify that the ECG rate and lead being monitored are displayed to the right of the ECG
parameter key.
• Verify that the ECG waveform is displayed.
3. Connect a waveform simulator to the pressure inputs.
4. Label and zero the channels.
• Verify that the numerics and waveforms are accurate.
5. Touch the RECORD key, then touch one of the flashing parameter keys.
• Verify that the recording is printed.
Optional Remote Alarm
If the monitor is configured to use the Remote Alarm interfaced into a nurse call station, plug the
cable into the Remote Alarm connector and initiate an alarm. The nurse call should respond.
Optional Wired Ethernet
If the monitor is configured to use the wired Ethernet, connect a 10BaseT cable from an active
Spacelabs Medical network into the rear panel connector. Press the SPECIAL FUNCTIONS key and
verify that other monitors appear in the Remote View submenu.
!
• If the monitor is properly configured and installed on a network, perform the
following test as described. The recordings will be printed at the central printer.