Service manual
UCW and Ultraview 1700 Monitors — Service Manual
4-2
To clean the exterior of monitors, modules, and cables:
• Prepare the cleaning solution according to the manufacturer’s instructions.
• Wet a clean cloth with the selected cleaning solution.
• Remove excess liquid from the cloth and squeeze dry.
• Wipe exposed surfaces of the equipment and cables.
• Remove any soap residue by gently wiping with a clean damp cloth.
• Wipe dry with a clean dry cloth.
Use only the following recommended cleaning solutions:
• Mild soap and water solution
• U.S. Pharmacopoeia (USP) green soap
• Sodium hypochlorite solution (1:10 dilution of household chlorine bleach in water)
• Phenolic germicidal detergent (1% aqueous solution)
• Glutaraldehyde (2.4%) (Cidex)
• Isopropyl alcohol (70% solution)
Tape adhesive can be removed with Spacelabs Medical adhesive tape remover pads
(P/N 392196-001).
Questions and concerns about cleaning issues should be directed to a Spacelabs Medical field
service engineer.
Filter and Fan (90385 Only)
To clean the fan and filter:
1. Remove the filter cover from the rear of the monitor using a small, flat head screwdriver to
gently pry off the cover.
2. Clean the filter by washing it with water and mild soap (or replace it).
3. Allow the filter to dry.
4. Reinstall the filter and the fan filter cover.
5. Verify that the fan is working properly and that it is drawing air into the display assembly.
Touchscreen (90385 Only)
The touchscreen does not have a separate shield over its face. Clean the screen with a soft cloth
moistened with either 70% alcohol or with soapy water.
!
• Over time, repeated use of a chlorine bleach solution may cause some colors to
fade.
!
• Depending upon the hospital environment, the air filter may need to be cleaned
more than once per year. It is recommended that, in the beginning, hospitals
inspect the air filter at monthly intervals until an appropriate cleaning cycle is
established.
!
• Do not allow liquid to enter the monitor.