Service manual
UCW and Ultraview 1700 Monitors — Service Manual
2-30
MONITOR CALIBRATION Key (90387)
The MONITOR CALIBRATION key enables the Biomed to access the Monitor Calibration menu to
make horizontal and vertical adjustments of the display. The position and size of the 90387 display is
a function of the external display device. The CRT POSITION key displays a border used to facilitate
the display adjustment.
Figure 2-13: 90387 Monitor Calibration menu
SYSTEM INFO Key
The SYSTEM INFO key displays an informational screen. Horizontal dividing lines above and below
this display separate it from other information, and four sections of data are presented, including:
• PCBA Information
• System Information
• Diagnostics
• Software Information
Touch PRINT to print the information displayed.
CHANGE PASSWORD Key
The CHANGE PASSWORD key enables you to change the case-sensitive password that allows
access to the Privileged Access menus.
To set or change the password:
1. Touch CHANGE PASSWORD to display the on-screen keyboard and the password
prompts.
2. Type your new password (nine characters maximum) at the New Password prompt. Touch
ENTER.
3. Enter this new password a second time at the Verify Password prompt. Touch ENTER.
SYSGEN Key
RESET MONITOR Key
Touch this key to reset the monitor after changing configuration settings, etc.
!
• Monitor configuration using the Sysgen feature is only accomplished by
Spacelabs Medical field service engineers.
BIOMED LEVEL - Calibrate Monitor Subsystems
CRT
POSITION
SYNC ON GREEN
ON
OFF
PRINT
SYSTEM INFO - DIGITAL