Service manual

Setup
2-29
Figure 2-12: 90385 Monitor Calibration menu
The Touchscreen Margin Adjustment/Settings menu provides preset settings to enable the
Biomed to tune a particular ring PCBA, IR lens, and I/O PCBA for optimal performance.
To adjust the CRT’s horizontal and vertical position (90385 only):
1. Touch the CRT POSITION key.
2. Verify that the CRT position adjustment keys appear (refer to Figure 2-12 on page 2-29).
3. Touch the adjustment keys as necessary to achieve the desired screen position and size.
The single arrow keys shift the raster in the indicated direction.
The double arrow keys expand (← →) or contract (→ ←) the raster image either horizontally or
vertically as indicated by the placement of the arrows ( or).
4. Touch the BORDER key to display a border around the display during adjustment.
WARNING:
Normally, there should be no reason to adjust the factory defaults. Improper
adjustments or calibrations can SEVERELY degrade touchscreen
performance.
!
If the display position is improperly set, the touchscreen will not align to the
display. If this condition occurs, shut off the monitor, enter two-finger diagnostics
(refer to
Extended Diagnostics on page 5-5) and set the screen position to mid-
range.
BIOMED LEVEL - Calibrate Monitor Subsystems
TOUCHSCREEN MARGIN
ADJUSTMENT
TOUCHSCREEN
CALIBRATION
BIOMED LEVEL - Touchscreen Margin Settings
DEFAULT
B
CRT
POSITION
BIOMED LEVEL - CRT Position
CDEA
BORDER
90385 only
SYNC ON GREEN
ON
OFF