UCW and Ultraview 1700 Monitors — Service Manual
2-28
Service Functions Menu
The Service Functions menu provides the operator with the capability to review system information
and calibrate the monitor’s display subsystem.
DISPLAY ERROR LOG Key
The DISPLAY ERROR LOG key enables you to display the monitor’s logged errors. Keys in this
menu enable you to page backward or forward through the error log, display the most recent page of
logged errors, clear the log, and print the log’s currently displayed page.
MONITOR CALIBRATION Key (90385)
The MONITOR CALIBRATION key enables the Biomed to access the Monitor Calibration menu in
order to make horizontal and vertical adjustments of the CRT screen’s display and to calibrate the
infrared touchscreen.
!
• Error logs are intended for use by Spacelabs Medical service personnel only and
do not necessarily indicate a need for monitor service.
!
• Touching SYNC ON GREEN for a monitor without that capacity may make the
screen unreadable. Be certain of the display capability before selecting this
option.
BIOMED LEVEL - Service Functions
MONITOR
CALIBRATION
DISPLAY
ERROR LOG
SYSTEM
CHANGE
PASSWORD
SYSGEN
RESET
MONITOR
INFO