Service manual

Setup
2-27
User Access Menu
The User Access menu provides privileged users access to enable certain functions for normal use.
The highlighted option is the one that is currently active. Touching another key changes the
selection.
The PATIENT TYPE ON/OFF key enables Adult/Neonate selection in the Monitor Setup menu.
The PARAMETER CONFIGURATION ON/OFF key displays the PARAMETER CONFIGURATION
key in the Monitor Configuration menu.
The RECORDING DURATION ON/OFF key controls user access to the recording duration feature.
When set to ON, the RECORDING DURATION key is displayed in the Recorder Configuration
menu.
UNITS OF MEASURE MENU Key
Touch the UNITS OF MEASURE key to select a unit of measure for the monitor:
The active option is highlighted. Touching another key changes the selection.
Tone Configuration Menu
The Tone Configuration menu enables the operator to configure the monitor’s alarm tones. The
active option is highlighted.
Touch CONFIGURABLE ALARM TONES to display the following menu.
This menu enables the operator to configure the monitor’s alarm tones to have different repetition
rates than are defined in the ISO standard.
USER ACCESS - Select user access-to functions
PATIENT TYPE
ON OFF
ON
PARAMETER CONFIGURATION
ON OFF
RECORDING DURATION
OFF
inches
cm
Restart monitor after selecting units of measurement
mmHg
kPa
lb
Kg
TONE CONFIGURATION - Restart the monitor after selecting tone configuration
CONFIGURABLE
ALARM TONES
ISO STANDARD
ALARM TONES
CONTINUOUS
ALARM TONES
Use the arrow keys to adjust the priority period of the selected alarm
HIGH
15s
MEDIUM
30s
LOW
30s
FACTORY
DEFAULTS