Service manual
UCW and Ultraview 1700 Monitors — Service Manual
2-26
RELAY TIMEOUT 0 SEC/10 SEC Key
The RELAY TIMEOUT key sets the number of seconds (0 or 10) that the alarm relay remains closed
following the end of an alarm condition.
FLASHING/STEADY ON Key
The FLASHING/STEADY ON key selects the mode of operation for the alarm relay. In Flashing
mode, the alarm relay contacts open and close according to the priority of the active alarm level. In
the Steady On mode, the relay contacts close if any alarms are active. They open only when all
alarm conditions cease.
ALARM LEVEL Key
The ALARM LEVEL key defines the level of alarms responded to by the monitor (activate alarm
relay): high, medium, or low.
Touching ALARM LEVEL displays the Alarm Level menu:
Touch HIGH to initiate only high-priority alarms only.
Touch MEDIUM to initiate medium- and high-priority alarms.
Touch LOW to initiate all alarm priorities (high, medium, low).
MINIMUM VOLUME ON/OFF Key
The MINIMUM VOLUME ON/OFF key locks (ON) the current alarm tone volumes as the new
minimum volumes or unlocks (OFF) the tone volume controls.
QRS/SPO2 TONE ENABLE Key
The QRS/SPO2 TONE ENABLE key defines whether the QRS or SpO
2
tones, when enabled within
the ECG or SPO2 menus, sound all the time (Always) or only when this monitor is in alarm (During
Alarm).
ALARM WATCH TONE Key
The ALARM WATCH TONE key defines whether the alarm watch notification tone sounds one to
three times whenever this monitor receives a new alarm watch multicast message.
MEDIUM
LOW
ALARM LEVEL - select minimum alarm priority level to trigger relay activation
HIGH