Service manual

Setup
2-25
To specify the serial port’s use:
1. From the Biomed menu, touch the SERIAL PORTS key.
2. Select a serial port key.
3. Verify that the SETTINGS and ASSIGNMENT keys appear.
4. Touch ASSIGNMENT and verify that the DIAGNOSTICS and DATA LOGGER keys appear.
5. Select the key that matches the application attached to this serial port (if diagnostic
messages are not desired during power-up, touch the DIAGNOSTICS key to disable ASCII
dumps to the serial port).
Alarm Setup Menu
The ALARM SETUP key enables you to configure the alarm features.
Touching this key displays the Alarm Setup menu:
REMOTE ACCESS ON/OFF Key
The REMOTE ACCESS ON/OFF key enables (ON) or disables (OFF) a remote monitor’s access to
this monitor’s alarm limits.
ALARM SUSPEND ON/OFF Key
The ALARM SUSPEND ON/OFF key enables (ON) or disables (OFF) operation of the monitor’s
ALM Suspend function.
TREND SUSPEND ON/OFF Key
When the TREND SUSPEND ON/OFF key is set to ON, trending is disabled when alarms are
suspended using the ALARM SUSPEND hard key. When this key is set to OFF, trending operation
becomes independent of the ALARM SUSPEND hard key.
ALARM RELAY Key
Touching ALARM RELAY displays the Alarm Relay Setup menu:
ALARM SETUP - Select parameter to change
REMOTE ACCESS
ON OFF
TREND SUSPEND
ON OFF
ALARM
RELAY
QRS/SPO2 TONE ENABLE
ALWAYS DURING ALARM
ALARM WATCH TONE
MORE
SINGLE TRIPLE
ALARM SUSPEND
ON OFF
FLASHING
STEADY ON
ALARM
LEVEL
RELAY TIMEOUT
0 SEC 10 SEC
ALARM RELAY SETUP - Select parameter