Service manual

UCW and Ultraview 1700 Monitors — Service Manual
2-20
Biomed Setup Menu
Network Setup Menu
The Network Setup menu enables you to define a subnet name, node ID, and bed name; view the
active nodes on the network; and select an appropriate subnet access level.
Touch NETWORK SETUP to display the following menu:
SUBNET Key
Touch SUBNET to display the Keyboard menu, enabling you to type in a subnet name (refer to
Keyboard Menu on page 2-19 for instructions on usage). This name can contain up to five
characters, letters, or numbers.
To enter a subnet name:
1. Touch SUBNET.
2. Type in the subnet name.
3. When you have entered the name correctly, touch ENTER to store the name in memory.
4. Reset the monitor so that the network can recognize the new name.
NODE ID Key
Touch NODE ID to display the pop-up keypad.
Each monitor on the network must have a unique node ID. The default node ID for an uninitialized
monitor is 1. You may change this ID to allow the monitor to function properly on the network. After
changing the ID, you must reset the monitor to enable the system to recognize the new node ID. All
monitors on a network must have a unique node ID.
To enter a node ID:
1. Touch NODE ID.
2. Enter a number between 1 and 64 (between 1 and 250 with monitors that support the
expanded network feature).
3. Touch ENTER on the keypad to place it into memory.
!
Additional keys may be present. Refer to the corresponding sections of this
manual or to the Ultraview Care Network Operations Manual (P/N 070-1001-xx,
located on CD-ROM 084-1101-xx) for configuration.
NODE
DIRECTORY
SUB
NET
NETWORK SETUP - Restart monitor after entering changes
NODE
ID
BED
NAME
SUBNET ACCESS
ON
OFF
TCP/IP
SETUP