Service manual

UCW and Ultraview 1700 Monitors — Service Manual
2-14
Ethernet LAN Installation
To install a 90385/90387 monitor onto an existing Spacelabs Medical Ethernet LAN, complete the
following (refer to
Figure 2-5 on page 2-9, Figure 2-6 on page 2-10, Table 1 on page 2-8, Table 2 on
page 2-8, and Table 3 on page 2-9).
1. Power OFF the monitor.
2. Attach one end of a 10BaseT Ethernet cable (C) to of the monitor.
3. Power ON the monitor and refer to System Configuration on page 2-14 to set up the monitor
for network operation.
4. Connect the other end of the Ethernet cable (C) from the monitor to the nearest network port.
System Configuration
The Biomed menu enables an authorized service representative to set user defaults for monitor
operation.
To access the Biomed menu:
1. Touch MONITOR SETUP.
2. Touch PRIVILEGED ACCESS to display the Keyboard menu (refer to Keyboard Menu on
page 2-19).
3. Enter the correct password (case-sensitive) to display the Biomed Monitor Setup menu
(refer to
Figure 2-9).
!
Detailed installation instructions for an Ethernet local area network (LAN) are
beyond the scope of this document.
CAUTION:
Only qualified personnel should attempt to install a monitor onto an
Ethernet LAN.
!
Do not connect the monitor to an Ethernet LAN prior to configuring the network
parameters. If it is incorrectly configured for LAN access, other units on the
network may be interrupted.
WARNING:
When 10Base5 hardware is used for the LAN, make sure the Ethernet wall
plate and the shield of the Ethernet connecting cable are bonded to the
hospital grounding system.
!
Based on your software version, your screen may differ from the following.