Service manual
UCW and Ultraview 1700 Monitors — Service Manual
2-10
Figure 2-6: 90387 cabling diagram
* The J9 COM port must be used to interface a serial touchscreen. Using J15 to interface the touchscreen is
not supported.
** Only one display with touchscreen can be attached to the 90387. A second display can only be driven for
video. Audio is not available.
CAUTION:
• To prevent accidental removal of the audio output connector, loop the cable
at least two times around the supplied strain relief, which is located at the
rear bottom corner of the 90387. Refer to
Figure 2-7.
• Examine any external display for strain relief capability and set it up per
manufacturer recommendations. Spacelabs 91415 and 91418 displays
incorporate the same strain relief indicated in
Figure 2-7. Any remote
display incorporating audio must have cable secured at both ends to assure
reliable audio at both monitor and display.
90387
90499
module
housing
Flexport
terminator
J7
J2
J14
J15
P1
J2
nurse
alert
serial I/O
Ethernet
A
E
C
G
DC power
supply
J16 P1
G
J
J18
J9
J10
H
E
K
serial I/O
*
display with
touchscreen
and audio
91416-B
display
J19
**
**
J7
J9
L
generic