Service manual
Setup
2-9
Figure 2-5: 90385 cabling diagram
* Video output is a repeater display and is not configurable.
** Serial output may also be used for external touchscreen input in some external display applications.
Cables (90387)
The cables and adapters available for making interconnections to the 90387 and other components
are listed in Table 2 on page 2-8 and Table 3 below. The use of these cables is illustrated
schematically in Figure 2-6 on page 2-10.
Refer to Parts on page 6-1 for additional cable part numbers.
CAUTION:
• Spacelabs Medical has tested and approved the cables listed in Table 3 for
use with the 90387. For continued EMI radiation compliance, use only these
cables.
Table 3: 90387 Cables and Adapters
Cable/
Adapter
P/N Description
G 012-0601-00 Cable, SDLC with Power Interface, 1 foot (0.305 m)
H 012-0593-00 Cable, Video
J 012-0555-00 Cable, Powered Flexport
K 012-0595-00 Cable, Audio
L 012-0620-00 Cable, Touchscreen/video Sync
monitor
nurse alert
Ethernet
serial I/O
**
DC
module housing
A
B, G
C
E
supply
power
ferrite beads
90491/90499
Video
*
F
90385
P1
SDLC