Service manual
Setup
2-7
Forced Air Fan — Small, quiet muffin fans rated at 100 cubic feet per minute (CFM) or
equivalent (to be provided by the institution) can be installed. In this case, provide one fan
for every central station or two system printers/module recorders that are installed in the
console.
An example console design is shown in Figure 2-4. This figure shows the minimum dimensions
required to satisfy the above requirements.
Figure 2-4: Minimum cavity sizes for console setup
!
• If a network printer is to be installed in the console, the power dissipation and
dimensional requirements will be different. Refer to the printer specifications for
further information.
!
• The dimensions shown in Figure 2-4 may not allow for appropriate clearance of
displays larger than 15 inches.
26.5”
(67.3 cm)
optional
air inlet
30”
(76.2 cm)
46.5”
(118.20 cm)
16.5”
(41.94 cm)
10×
cable access
hole
air inlet
AC power
removable back
exhaust fan
with louvers
optional DC power
supply mount using wall
mount
DC power supply