Service manual
UCW and Ultraview 1700 Monitors — Service Manual
2-6
Figure 2-3: Pedestal mount
Console Mounting (90387)
The physical layout and other console characteristics depend on the requirements of the hospital or
other institution. It is the responsibility of the institution to define its specific console needs and to
design and construct the console according to its specifications.
For this reason, Spacelabs Medical does not supply a pre-fabricated console assembly. Instead, this
section provides the information necessary to create a custom console design. In designing the
console, the following considerations must be kept in mind.
• Access
The rear of the console must be designed to allow for easy access. The dimensions of the rear
access door must allow the 90387 and display monitor to be easily installed and removed.
• Cooling
The console cabinet must be designed to allow sufficient air flow for natural convection.
If the console cabinet is enclosed, one of two cooling methods can be used:
Central Air Conditioning — The console can be connected directly to an air conditioning
cooling duct. The maximum ambient air temperature should be limited to approximately
32°
C, and vented from the console at the highest exit.
!
• The 90387 does not contain a fan.