The wall mounting hardware consists of:
Wall Track — The wall track is mounted vertically on the wall with the mounting arm installed in
the track. The mounting arm can be slid up and down in the wall track to the desired height. Wall
tracks are available in multiple lengths.
Wall Track End Caps — The wall track end caps are installed on the top and bottom of the wall
track to give the track a finished appearance and to prevent the equipment from sliding off the
end of the mounting track.
Mounting Arm — The mounting arm projects horizontally out from the wall track and holds the
display assembly and module housings. Two types of arms are available: one for a monitor-only
installation; the other for installation of the monitor and module housing(s).
Mounting Screws — Mounting screws are provided to install the end caps and to secure the
mounting arm at the desired vertical position.
Mounting Considerations
Power supply should be placed no closer than three inches (7.62 cm) to the remote housings to
allow sufficient air flow. Install a minimum of four AC outlets.
Pedestal Mounting (90385)
To attach the pedestal mount to the display assembly refer to Figure 2-3.
1. Verify that the pedestal’s six rubber feet and four retaining studs with retaining rings have
not been dislodged during shipping.
2. Carefully place the display on its side. A cushion should be used under the display to avoid
scratching the display case.
3. Rotate the display’s round base (if necessary) so that its tick mark points toward the front of
the unit.
4. Line up the four retaining studs on the pedestal with the four holes in the display’s round
base.
5. Push the pedestal onto the round base and secure it by turning each of the four retaining
studs 90 degrees clockwise with a straight blade screwdriver. The studs will lock in place
and can no longer be turned.
6. Carefully return the display to its normal upright position.
!
The monitor, monitor wall plate, and AC outlet must be located on the same side