Appendix A: Technical Specifications - EMC Considerations A-5
Guidance and Manufacturers Declaration – Emissions
Equipment and Systems that are NOT
Life-supporting
Guidance and Manufacturer’s Declaration – Emissions
The ClearCount SmartSponge Flex Model A04 is intended for use in the electromagnetic environment
specified below. The customer or user of the ClearCount SmartSponge Flex Model A04 should ensure that it
is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment – Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3Vrms
3V/m
Portable and mobile communications equipment
should be separated from the ClearCount
SmartSponge Flex Model A04 by no less than the
distances calculated/listed below:
D=(3.5/3)(Sqrt P)
D=(3.5/3)(Sqrt P)
80 to 800 MHz
D=(7/3)(Sqrt P)
800 MHz to 2.5 GHz
where P is the max power in watts and D is the
recommended separation distance in meters.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
,
should be less than the compliance level in each
frequency range
b
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
.