User's Manual
Table Of Contents
- Preface
- Chapter 1: System Description
- Chapter 2: System Interface
- Chapter 3: Preparing for a New Case
- Chapter 4: System Operation
- Chapter 5: System Settings
- Chapter 6: Review Case Data
- Chapter 7: Cleaning and Maintenance
- Chapter 8: Troubleshooting
- Appendix A: Technical Specifications
- SmartSponge® Flex System Dimensions
- Power Requirements
- Environmental Conditions
- SmartSponge Flex Sponges and Towels
- MRI Information
- EMC Considerations
- Guidance and Manufacturer’s Declaration – Emissions
- All Equipment and Systems
- Guidance and Manufacturer’s Declaration – Immunity
- All Equipment and Systems
- Guidance and Manufacturer’s Declaration – Emissions
- Equipment and Systems that are NOT Life-supporting
- Recommended Separation Distances between portable and mobile RF Communications equipment and the ClearCount SmartSponge Flex Model A04
- Equipment and Systems that are NOT Life-supporting
- Device Label
Chapter 8: Troubleshooting - System Alerts 8-3
•
•
•
•
•
•
Basin Limit Reached
Wand Not Functioning
CAUSE:
There are over
20
sponges in the Count Out
Basin.
ACTION:
Remove sponges or discard the full bag and replace with
a new liner - sponge counts will not change.
Each additional sponge or group of sponges discarded
into the Out Scan after its capacity is already at
20
will
trigger this System Alert.
CAUSE:
The SmartWand is either not connected,
detuned due to its proximity to metal (still in
the wand hanger), or experiencing a hardware
malfunction.
ACTION:
Ensure that the SmartWand is properly plugged into the
system and that it is adequately spaced from metal (this
includes the wand hanger).