User's Manual
Table Of Contents
- Preface
- Chapter 1: System Description
- Chapter 2: System Interface
- Chapter 3: Preparing for a New Case
- Chapter 4: System Operation
- Chapter 5: System Settings
- Chapter 6: Review Case Data
- Chapter 7: Cleaning and Maintenance
- Chapter 8: Troubleshooting
- Appendix A: Technical Specifications
- SmartSponge® Flex System Dimensions
- Power Requirements
- Environmental Conditions
- SmartSponge Flex Sponges and Towels
- MRI Information
- EMC Considerations
- Guidance and Manufacturer’s Declaration – Emissions
- All Equipment and Systems
- Guidance and Manufacturer’s Declaration – Immunity
- All Equipment and Systems
- Guidance and Manufacturer’s Declaration – Emissions
- Equipment and Systems that are NOT Life-supporting
- Recommended Separation Distances between portable and mobile RF Communications equipment and the ClearCount SmartSponge Flex Model A04
- Equipment and Systems that are NOT Life-supporting
- Device Label
5-2
Chapter 5: System Settings
-
Edit Settings Mode
•
•
•
•
•
•
Edit Settings Mode
The SmartSponge Flex allows you to adjust some system settings and parameters to aid your use of the system.
Press the Edit Settings icon from the Case Selection screen to enter the Edit Settings mode. This will trigger
the Override screen. An Override Card will need to be placed on the In Scanner of the system to proceed to the
Edit Settings mode. The system will enter the Edit Settings mode after the Override Card is scanned.
Figure 5-1 Enter Edit Settings Mode
Swipe the
Override Card
across the
In Scanner
.
Press the
Edit Settings
Icon
to trigger the
OVERRIDE
screen.