User's Manual
Table Of Contents
- Preface
- Chapter 1: System Description
- Chapter 2: System Interface
- Chapter 3: Preparing for a New Case
- Chapter 4: System Operation
- Chapter 5: System Settings
- Chapter 6: Review Case Data
- Chapter 7: Cleaning and Maintenance
- Chapter 8: Troubleshooting
- Appendix A: Technical Specifications
- SmartSponge® Flex System Dimensions
- Power Requirements
- Environmental Conditions
- SmartSponge Flex Sponges and Towels
- MRI Information
- EMC Considerations
- Guidance and Manufacturer’s Declaration – Emissions
- All Equipment and Systems
- Guidance and Manufacturer’s Declaration – Immunity
- All Equipment and Systems
- Guidance and Manufacturer’s Declaration – Emissions
- Equipment and Systems that are NOT Life-supporting
- Recommended Separation Distances between portable and mobile RF Communications equipment and the ClearCount SmartSponge Flex Model A04
- Equipment and Systems that are NOT Life-supporting
- Device Label
4-30
Chapter 4: System Operation
•
•
•
•
•
•
Restoring Power
In the event of a power failure, move the power cord from a standard wall outlet to a red emergency powered
outlet if available. Ensure the On/Off switch remains in the On
( | )
position. When the system restarts it will
continue with the current case and advance to Standby mode. The screen will notify you that the case is in
progress. Press the appropriate button to enter the desired mode. The system will resume counting exactly
where it left off. Sponge counts are maintained in the event of power loss and are resumed when power returns.
If the power cord is accidentally unplugged during use from either the wall outlet or power entry module,
replace the power cord to the wall outlet and/or the power entry module. The system will automatically resume
back to the Standby screen as long as the On/Off switch is still in the On
( | )
position. Press the appropriate
button to enter the desired mode. All sponge counts are stored in the system’s database whenever there is a loss
of power. Sponge counts are resumed upon the return of power.