User's Manual
Table Of Contents
- Preface
- Chapter 1: System Description
- Chapter 2: System Interface
- Chapter 3: Preparing for a New Case
- Chapter 4: System Operation
- Chapter 5: System Settings
- Chapter 6: Review Case Data
- Chapter 7: Cleaning and Maintenance
- Chapter 8: Troubleshooting
- Appendix A: Technical Specifications
- SmartSponge® Flex System Dimensions
- Power Requirements
- Environmental Conditions
- SmartSponge Flex Sponges and Towels
- MRI Information
- EMC Considerations
- Guidance and Manufacturer’s Declaration – Emissions
- All Equipment and Systems
- Guidance and Manufacturer’s Declaration – Immunity
- All Equipment and Systems
- Guidance and Manufacturer’s Declaration – Emissions
- Equipment and Systems that are NOT Life-supporting
- Recommended Separation Distances between portable and mobile RF Communications equipment and the ClearCount SmartSponge Flex Model A04
- Equipment and Systems that are NOT Life-supporting
- Device Label
Chapter 4: System Operation 4-29
•
•
•
•
•
•
Ending a case with “COUNTS NOT EQUAL”:
Step 1 Press the
CASE SUMMARY
button located at the bottom of the screen.
Case Summary mode can be entered from any other mode. If the counts are not reconciled, the Case
Summary screen displays “
COUNTS NOT EQUAL
”. See
Figure 4-20
.
Step 2 With “COUNTS NOT EQUAL” displayed at the top of the Case Summary screen, press the
OVER RIDE
button located at the bottom of the screen.
The “
SCAN ADMIN CARD TO PROCEED
” message will now be displayed on the OVERRIDE
screen. (Press BACK if you desire to return to the previous screen)
Step 3 The person responsible for the Override Card will need to present the card. Scan the Override
Card by placing it onto the In Scanner and holding it there until an audible alert is heard and the
system returns to the Case Summary screen.
“
VERIFIED BY ADMIN
” will now appear in the count status of the report. See
Figure 4-20
.
Step 4 Enter the Case ID into the patient’s record from the top line of the report.
Sponges may be intentionally withheld from the Count Out Basin for procedural or clinical reasons.
Alert the OR manager, and note this on the patient’s record along with the Case ID.
Step 5 Press the END CASE button to end the case.
The system will save the case data and return to Case Selection mode.
Step 6 Remove the System Cover that contains the discarded sponges from the Count Out Basin.
Dispose of the bagged items according to the standard protocol for your hospital.
Step 7 Clean the SmartSponge Flex according to the procedure in Chapter 7 before entering it into the
next surgical case.