User's Manual

Table Of Contents
Chapter 4: System Operation 4-17
Scanning a Patient for Retained Items
The SmartWand allows you to scan patients for retained ClearCount sponges and towels at any point during
surgery. Remove wand from holder then press the WAND button to enter Wand mode. It is important to note
that an item found by the wand is not counted out. An item can only be counted out when it is presented to the
Count Out Basin and the system is in Count Out mode. The SmartWand performs best when passed over the
patient in a slow, steady fashion, no faster than 0.2 m/second (approximately 7 inches/second). Maintain a
distance of 1 to 2 inches above the patient. While a tagged item may be found on the first pass of the wand, it
is typical to make several passes at varying inclinations to locate an item within a body. On a typical patient,
each scan pass should take approximately 5 seconds to complete. See
Scanning Procedure
.
Figure 4-13 Wand Mode Screen
To clear the
detected sponges
and SmartTag
detection from the
screen, press the
RESET
button.
Press the
WAND
button from any
mode to enter
Wand Mode.
When sponges are
detected, they will
be displayed by
type and quantity.
When the
SmartTag is
detected, the
green check
mark appears.
SmartTag and 2 4x4 detected