Chapter 4: System Operation
Warning!
Holding items that
have not been
scanned in too close to the Count Out
Basin while in Count Out mode may result in the items unintentionally being
detected prior to use. Follow the on screen prompts and decline acceptance if
the sponges were inadvertently held too close to the Count Out Basin.
Holding items that
have already been
scanned in too close to the Count Out
Basin while in Count Out mode may result in the items unintentionally being
counted out prior to use. Dispose any items into the basin that have been
scanned in and then counted out (detected) by the Count Out Basin.
Notes
Make sure “COUNT IN” is displayed in the upper right of the display
when scanning packs into the surgical case.
If two or more packs are detected by the In Scanner at the same time the
system will display the “Multiple Packs Detected” alert. Remove the
packs and re-scan one at a time.
If a pack that has already been scanned in is presented to the In Scanner
to be scanned in again, the message
Pack Already Scanned” will
appear. Open and use pack as per normal operation.
If the alert “Discard Pack” is displayed, throw out the defective pack
and replace it with a new pack.
If after a pack is scanned, no tone is heard and zero sponges are added to
the count, throw out the defective pack and replace it with a new pack.
Do not rest sponge packs or any other items on or near the In Scanner.