User's Manual
Table Of Contents
- Preface
- Chapter 1: System Description
- Chapter 2: System Interface
- Chapter 3: Preparing for a New Case
- Chapter 4: System Operation
- Chapter 5: System Settings
- Chapter 6: Review Case Data
- Chapter 7: Cleaning and Maintenance
- Chapter 8: Troubleshooting
- Appendix A: Technical Specifications
- SmartSponge® Flex System Dimensions
- Power Requirements
- Environmental Conditions
- SmartSponge Flex Sponges and Towels
- MRI Information
- EMC Considerations
- Guidance and Manufacturer’s Declaration – Emissions
- All Equipment and Systems
- Guidance and Manufacturer’s Declaration – Immunity
- All Equipment and Systems
- Guidance and Manufacturer’s Declaration – Emissions
- Equipment and Systems that are NOT Life-supporting
- Recommended Separation Distances between portable and mobile RF Communications equipment and the ClearCount SmartSponge Flex Model A04
- Equipment and Systems that are NOT Life-supporting
- Device Label
Chapter 4: System Operation 4-3
•
•
•
•
•
•
Standby Mode
The SmartSponge Flex is designed to not perform any scanning function until you specifically select to do so.
Therefore, the system will enter Standby mode when you select to begin a new case. While in Standby mode,
all antennas are off.
The system can remain in Standby mode for as long as necessary while you prepare for surgery. When you are
ready, you can navigate to any of the following modes:
•
Count In - Activates the In Scanner for scanning SmartSponges into the case.
•
Count Out - Activates the Count Out Basin for counting SmartSponges out of the case.
•
Wand - Activates the SmartWand for detecting SmartSponges.
•
Case Summary - Shows the sponge counts and case information for the current surgical case.
Figure 4-2 Standby Mode Screen