User Manual

Principles of Operation
5 Channel Calibrator and SGDM System
9
2.2.3 Sample Gas Distribution
1) Cease the flow of any calibration gases. Allow a few seconds for them
to vent from the systems if necessary.
2) Press the green “CAL MODE” button on the SGDM (2-4) to enter
Sample Mode. Be sure that it is NOT lit up.
3) Using the SGDM’s flowmeters (2-1) adjust the flow of sample gas to
the analyzers.
2.2.4 Control via a Remote Device
1) Set the calibration output selector
(4) to “A – AUTO”.
2) Set the calibration gas output selector knob
(3) to “AUTO”.
3) Insert the 7 pin Amphenol plug from the remote device into the
Amphenol input
(9) on the back of the 5 Channel Calibrator.
4) Insert the Amphenol from the remote device into the Amphenol input
(2-9) of the SGDM.
5) Plug in devices if you have not already done so.
6) Using the flow meters (1, 2-1), adjust the flow rate.
7) The input and output of the system can now be controlled with a
remote device.
NOTE: The flow rates of the devices cannot be controlled via a remote
device. Any changes must be performed manually.