User's Manual
Table Of Contents
- About This Manual
- About the Clarius Ultrasound Scanner
- Using the Clarius Ultrasound Scanner
- Accessories
- Cleaning & Disinfecting
- Safety
- References
- Measurement Accuracy Tables
- Acoustic Output Tables
- Clarius Scanner C3 HD3: B-Mode
- Clarius Scanner C3 HD3: Color Doppler Mode
- Clarius Scanner C3 HD3: M-Mode
- Clarius Scanner C3 HD3: PW Doppler Mode
- Clarius Scanner C7 HD3: B-Mode
- Clarius Scanner C7 HD3: Color Doppler Mode
- Clarius Scanner C7 HD3: M-Mode
- Clarius Scanner C7 HD3: PW Doppler Mode
- Clarius Scanner EC7 HD3: B-Mode
- Clarius Scanner EC7 HD3: Color Doppler Mode
- Clarius Scanner EC7 HD3: M-Mode
- Clarius Scanner EC7 HD3: PW Doppler Mode
- Clarius Scanner L7 HD3: B-Mode
- Clarius Scanner L7 HD3: Color Doppler Mode
- Clarius Scanner L7 HD3: M-Mode
- Clarius Scanner L7 HD3: Needle Enhance B-Mode
- Clarius Scanner L7 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L7 HD3: PW Doppler Mode
- Clarius Scanner L15 HD3: B-Mode
- Clarius Scanner L15 HD3: Color Doppler Mode
- Clarius Scanner L15 HD3: M-Mode
- Clarius Scanner L15 HD3: Needle Enhance B-Mode
- Clarius Scanner L15 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L15 HD3: PW Doppler Mode
- Clarius Scanner L20 HD3: B-Mode
- Clarius Scanner L20 HD3: Color Doppler Mode
- Clarius Scanner L20 HD3: Ocular (Ophthalmic) B-Mode
- Clarius Scanner L20 HD3: M-Mode
- Clarius Scanner L20 HD3: Needle Enhance B-Mode
- Clarius Scanner L20 HD3: PW Doppler Mode
- Clarius Scanner PA HD3: B-Mode
- Clarius Scanner PA HD3: Color Doppler Mode
- Clarius Scanner PA HD3: M-Mode
- Clarius Scanner PA HD3: PW Doppler Mode
- Clarius Scanner PA HD3: Transcranial B-Mode
- Clarius Scanner PA HD3: Transcranial Color Doppler Mode
- Clarius Scanner PA HD3: Transcranial M-Mode
- Clarius Scanner PA HD3: Transcranial PW Doppler Mode
- Revision History
Clarius Ultrasound Scanner - HD3 Scanners
revision 1 80
Clarius Scanner PA HD3
CLARIUS SCANNER PA HD3: B-MODE
Measurement Tolerance Range
Axial Distance < ± 2% 0 – 30.5 cm
Lateral Distance < ± 5% 0 – 30.5 cm
CLARIUS SCANNER PA HD3: M-MODE
Measurement Tolerance Range
Distance < ± 2% 0 – 30.5 cm
Time < ± 2% minimum = 0 ms
maximum = variable
a
a. Range depends on the viewing device used and the amount
of spectrum that can fit on the device's screen.
Heart Rate < ± 2% minimum = ≤ 1 beat
maximum = variable
a
CLARIUS SCANNER PA HD3: PW-MODE
Measurement Tolerance Range
Velocity < ± 2% 0 – 308.0 cm/s
Time < ± 2% minimum = 0 ms
maximum = variable
a
a. Range depends on the viewing device used and the amount
of spectrum that can fit on the device's screen.
Heart Rate < ± 2% minimum = ≤ 1 beat
maximum = variable
a
CLARIUS SCANNER PA HD3: DOPPLER SENSITIVITY
Sensitivity Depth (cm) Flow (ml/sec)
PW Doppler flow sensitivity at depth 5.0 0.1
CFI Doppler flow sensitivity at depth 5.0 0.1
PDI Doppler flow sensitivity at depth 5.0 0.1
PW Doppler depth sensitivity 9.4 –
CFI Doppler depth sensitivity 9.7 –
PDI Doppler depth sensitivity 10.2 –