Clarius Ultrasound Scanner - HD3 Scanners The Clarius Ultrasound Scanner
revision 1 65
To reach an operating temperature of 20°C (68°F), the Clarius Scanner HD3 requires
approximately 30 minutes to:
• Warm up from a storage temperature of -20°C (-4°F).
• Cool down from a storage temperature of 50°C (122°F).
Maximum scanner surface temperatures are:
• C3 HD3 = 35.5°C
• C7 HD3 = 35.5°C
• EC7 HD3 = 35.5°C
• L7 HD3 = 39.4°C
• L15 HD3 = 39.4°C
• L20 HD3 = 39.4°C
• PA HD3 = 39.4°C
If the scanner reaches its maximum surface temperature, it automatically shuts down.
To limit surface heating in the event of a single fault condition, the EC7 HD3 scanners
automatically shut down.
For information on storage temperatures see Storing Scanners on page 29.
Environmental Specifications
This icon, when blue, indicates that the scanner is cool. When this icon is red, it
indicates that the scanner is warm.
SCANNERS C3 HD3, C7 HD3, EC7 HD3, L7 HD3, L15 HD3, L20 HD3, & PA HD3
Operating Limits
Transient Operating Conditions
a
a. The conditions under which the scanner can operate for at least 20 minutes immediately
after being removed from an environment of 20°C (60°F), by combining the Fan, B-
Mode, and Eco Mode (extends scanner time at reduced frame rate).
Storage& Transportation Limits
Temperature C (32°F) to 40°C (104°F) -20°C (-4°F) to 35°C (95°F) -20°C (-4°F) to 50°C (122°F)
Humidity 15% to 95% RH 15% to 95% RH 0% to 95% RH
Pressure 620 hPa to 1060 hPa n/a n/a